Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 20
4.7.3 Download other parameters into volatile memory
4.7.3.1 For Non-digitally Compensated 9400 Sensors without Non-volatile Memory
or Non-9400 Sensors
A user-defined date can be downloaded into volatile memory using the ‘v
command with coefficient index 07. A manufacturing reference or serial
number can be downloaded into volatile memory using the ‘v’ command
with coefficient index 09 (format=5). A transducer full-scale range code
can be downloaded into volatile memory using the ‘v’ command with
coefficient index 0A. User-defined range codes can be assigned to values
128 and up (format=5). (See User’s Manual, Appendix E for standard
Range Codes.)
4.7.3.2 For Non-digitally Compensated 9400 Sensors with Non-volatile Memory
For non-digitally compensated 9400 sensors with non-volatile memory,
the same commands apply as for non-9400 generic” (third party) sensors.
The difference is that for non-digitally compensated 9400 sensors, upon
command, parameters are immediately stored into the sensor’s
non-volatile memory.
4.7.4 Set parameters to factory defaults
4.7.4.1 For Non-digitally Compensated 9400 Sensors Without Non-volatile Memory
or Non-9400 Sensors
All of the above parameters can be set to factory defaults using the ‘w
command with index 25. For each third-party type sensor attached, this
command sets the volatile copy of each parameter back to the factory
default value as follows:
Offset term = 0.0
Gain term = 1.0
c0 = 0.0
c1 = 1.0
c2 = 0.0
c3 = 0.0
user-defined date = 0
manufacturing reference number = 0
full-scale range code = 0
front end gain = 0 (unity).