bbbb (hex)
data selected for inclusion in each stream packet
0001 **
Enable Valve Position Status (reserved for future use)
0002
Enable DH Temperature Status (see bit map below)
0010
Enable Pressure EU Data (default if “05” never executed
after “00”)
0020
Enable Pressure A/D Counts
0040
Enable Pressure Voltages
0080
Enable DH Temperature EU Data (degrees C)
0100
Enable DH Temperature A/D Counts
0200
Enable DH Temperature Voltages
Any DH Temperature Status datum is delivered as a two-byte binary bit map
(16-bit, big endian) with each bit representing the status of DH #16 through DH #1
respectively. A bit value of 0 (zero) indicates the DH is operating within its
specified limits. A value of 1 (one) indicates the DH is outside its specified
limits.
Bit #
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Chan #
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Binary
1
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
Hex
8
0
0
1
The above example indicates that Channels 1 and 16 are operating outside the
specified temperature limits.
NOTE: Selecting too many other data groups will compromise module
performance.